Failure of VariAx® Distal Radial Locking Plate: Fractographic Analysis and Report of Two Cases
DOI:
https://doi.org/10.12974/2313-0954.2014.01.02.8Keywords:
Distal radius, osteotomy, locking plate, implant failureAbstract
We describe two cases of atraumatic fracture of VariAx® distal radial locking plate following its use in distal radial osteotomies. The plates fractured through an unused screw hole at the level of the osteotomy. Scanning electron microscopy was used to evaluate the fracture surfaces. This confirmed that there had been fatigue failure and did not find evidence of abnormal composition or structure of the metal. These cases are important because they highlight that this new and innovative implant can be prone to failure, and demonstrate the requirement for adequate surgical technique and metalwork protection following surgery.
References
Gautier E, Sommer C. Guidelines for clinical application of the LCP. Injury 2003; 34 Suppl 2: B63-76. http://dx.doi.org/10.1016/j.injury.2003.09.026
Imade S, Matsuura Y, Miyamoto W, H Nishi, Y Uchio. Breakage of volar locking compression plate in distal radial fracture. Injury Extra. 2009; 40(4): 77-80. http://dx.doi.org/10.1016/j.injury.2003.09.026
Tolat AR, Amis A, Crofton S, Sinha J. Failure of humeral fracture fixation plate in a young patient using the Philos system: case report. J Shoulder Elbow Surg. 2006; 15(6): e44-7. http://dx.doi.org/10.1016/j.jse.2006.01.014
Cooper EO, Segalman KA, Parks BG, Sharma KM, Nguyen A. Biomechanical stability of a volar locking-screw plate versus fragment-specific fixation in a distal radius fracture model. Am J Orthop. 2007; 36(4): E46-9.
Arora R, Lutz M, Fritz D, Zimmerman R, Oberladstatter J, Gabl M. Palmar locking plate for treatment of unstable dorsal dislocated distal radius fractures. Arch Orthop Trauma Surg. 2005; 125(6): 399-404. http://dx.doi.org/10.1007/s00402-005-0820-8
Koshimune M, Kamano M, Takamatsu K, Ohashi H. A randomized comparison of locking and non-locking palmar plating for unstable Colles’ fractures in the elderly. J Hand Surg Br. 2005; 30(5): 499-503. http://dx.doi.org/10.1016/j.jhsb.2005.04.018
Smith DW, Henry MH. Volar fixed-angle plating of the distal radius. J Am Acad Orthop Surg. 2005; 13(1): 28-36.
Tsangarakis N. Fatigue failure of an orthotropic plate with a circular hole. Journal of Composite Materials 1984; 18(1): 47- 57. http://dx.doi.org/10.1177/002199838401800104
Pilkey WD. Peterson’s stress concentration factors, 2nd edition. New York, J Wiley. 1997. http://dx.doi.org/10.1002/9780470172674